IOTA RGA Specifications
Quadrupole Analyser
Mass range: 1 – 100, 1 – 200, 1 – 300 amu Filaments: Twin Thoriated Iridium Filament Degas: Fully programmable filament degas facility Mass Filter: Single 6.3mm diameter Source Type: Open ultra high vacuum source Minimum detectable partial pressure: 5x10-13mbar Sensitivity: 5x10-4A torr-1 Faraday Operating temperature: 15 – 100 ºC Bakeout Temperature: 250 ºC maximum
Electronics Control Unit (ECU)
Length 125 mm Height 85 mm Width 105 mm Power +24 VDC at 2A Status 2 bi-colour led's Stand-alone operation (option) Universal mains power supply included
Connection Options
RS232 with USB converter & software (standard) Ethernet (option)
User I/O
2 off digital input (TTL) shared with open collector outputs 3 off open collector O/P (60v 1A total) shared with above 3 off DAC 16 BIT (0-10v Output) 2 off ADC 23 BIT (-10v to +10v Input) Aux Synchronous serial bus for I/O expansion Note: If digital external interlock facility is used then 1 of the two user inputs is lost to this function.
Standard Software
All standard scan types Multi-head, network operation Remote operation over intranet, internet etc. Firebird database server for storage Stand alone data review Powerful templates to customise scan presentation Extensive scheduling capability Extensive gas library

|
IOTA RGA Systems |
Description |
Detector |
Order Code |
|
100 AMU Single Detector |
Faraday only |
ZIQ100F |
|
200 AMU Single Detector |
Faraday only |
ZIQ200F |
|
100 AMU Dual Detector |
Faraday and multiplier |
ZIQ100D |
|
200 AMU Dual Detector |
Faraday and multiplier |
ZIQ200D |
|
300 AMU Dual Detector |
Faraday and multiplier |
ZIQ300D |
Contact us for further information on the IOTA RGA or to obtain trial software or a Live Access demonstration. |