Home Products Contact Us About Us Exhibitions Downloads Catalogue Careers

 

R3000 XPS/UPS

R3000This XPS/UPS model is equipped with XPS, UPS electronics and lens tables. It features overview angular modes and a transmission mode for high throughput measurements. Kinetic energies from 0.5 eV – 1500 eV are measured with a selection of slits and wide range of pass energies from 2-200 . Guaranteed energy resolution <3 meV at 20 eV kinetic energy.



Performance:

• Guaranteed energy resolution: < 3 meV @ 2 eV pass energy at 20 eV kinetic energy
• Pass energies: 2, 5, 10, 20, 50, 100, 200 eV
• Transmission mode lens acceptance: +/- 15 degrees
• Angular mode: ±5°, ±10°
• Kinetic energy range: 0.5 eV – 1500 eV
• Maximum energy window(single-shot): 24 eV, 12% of pass energy
• 45 mm working distance
• 40 mm diameter MCP detector system
• Bakeable to 150 degrees C

Included:

• Scienta R3000 mechanics and XPS/UPS electronics
• Computer system
• Windows based acquisition software: SES (analyser control software)
• Free upgrades of the SES-software for 10 years.
• Extremely stable and low noise electronics
• Strictly non-magnetic materials used in the interior
• Fix mode, 12 % of the pass energy
• Swept mode
• Angular mode

VG Scienta digital CCD-detector advantage:

• 600 energy channels available simultaneously
• 400 angular channels available simultaneously
• Total noise counts for all energy channels: < 2 cps
• Average noise per channel: < 0.01 cps
• Available dynamic range per pixel: > 7 bit
• Detector information read-out rate: 400 Mbit/s

The VG Scienta R3000 unit is a highly flexible analysis tool is optimized for quick PES measurements and band mapping. It features a modern fire wire CCD camera detection system for high data transfer and ease of upgrade. The high voltage electronics provides ultra stable conditions for reliable accurate measurements. The unit can be operated in “Quick Mode” where a spectrum can be recorded within seconds by taking a snap shot of the detector image, covering 12% of the pass energy. The analyzer is housed in a state of the art mu-metal vessel which gives a compact footprint.

The VG Scienta R3000 is a 135 mm mean radius hemispherical electron analyzer with wide angle acceptance lens for high troughput measurements. The analyzer is equipped with six interchangeable slits for optimal balance between best intensity and best energy resolution. The standard slit selection is: 0.2c, 0.4c, 0.8c, 1.3s, 3s and 0.2s where s=straight slit and c=curved slit.

The analyzer is also equipped with a real time read-out and electronic analyzer lens X,Y deflectors for diagnostics and experiment optimization purposes.  A 2-D low noise digital CCD-MCP detector system (1300x1000 pixel) with a noise level of < 0.01 cps/channel with USB connection to instrument computer is included.

DOWNLOAD DATA SHEET >>